Abstract:
Nanoindentation is derived from the classical hardness test but is carried out on a much smaller scale. It can be used to determine the hardness of thin layers as well as material properties such as elasticity, stiffness, plasticity, and tensile strength, or fracture toughness of small objects and microsystems in fi elds such as biotechnology. These measurements involve applying a small force to a sample using a sharp probe and measuring the resultant penetration depth. The measured value is used to calculate the contact area and hence the particular property of the sample material. Both the method of force application and the geometry of the indentation tip can be adjusted to suit the particular application.
Hardness, Elasticity and Tensile Strength Tests
The system can be used to slowly increase the load over a defined period of time ('quasistatic indentation'), or to perform a dynamic test where the tip is accelerated towards the surface, depositing a large amount of energy on contact (‘nano-impact'). The wear of materials can be assessed by moving the sample slowly to produce a scratch while applying either a constant or ramped load as the sample moves along. The frictional properties of the surface are measured simultaneously. The system applies the force using electromagnetic actuation. The NanoTest from Micro Materials Ltd. is suitable for use with a wide range of materials as it can apply forces of between 30 nN and 500 mN depending on the operating mode, and it can measure penetration depths of between 0.1 nm and 50 μm. This task is carried out by a high-resolution capacitive sensor, the PISeca sensor from PI. The capacitive sensor measures the penetration depth of the tip as a function of time. In static operation, the coil current and thus the load is measured at the same time, enabling a load-displacement diagram to be compiled. In dynamic operation, the load is derived from the effective acceleration of the loading head.
Scanning Probe Microscopy Optimizes the Evaluation
The indentation method employed by the NanoTest can be combined with an imaging technique to investigate the sample surface pre and post indentation. The indentation tip is moved row-by-row in a defined grid across the surface of the sample, that is, the sam ple surface is scanned. The displacement of the tip is recorded point-by-point and used to generate a three-dimensional image of the sample surface which can be evaluated with the appropriate software - also developed by Micro Materials.This combination of indenter and scanning probe microscope thus provides a quick method for assessing the sample surface. It is also possible to perform measurements at different locations without a great deal of effort.A P-527 piezo-based nanopositioning system from PI provides the motions in the X and Y axes which are required for the scan. Its 200 x 200 μm scanning range allows accurate statements to be made about the properties of the contact area and its surroundings. Thanks to the positioning accuracy of less than 3 nanometers a high positional resolution for the scan is guaranteed. This means, critical loads are accurately assigned to a defi ned position which in turn allows conclusions to be drawn about the local material properties, and the high number of closely packed measuring points make it possible to investigate very small samples.
Nanometrology: Capacitive Position Sensors
he absolute measuring capacitive sensors operate with resolutions down to the sub-nanometer range, and provide depth information without coming into contact with the sample. They operate without wear and hysteresis, and do not affect the application itself.
A simple physical principle is at the heart of capacitive metrology: When a voltage is applied, a homogenous electric fi eld is produced between the plates of a capacitor. When the gap between the plates changes, the output signal from the metrology electronics changes proportionally. Capacitive position sensors are available in two different basic designs, i.e. as one-electrode or two-electrode systems. The NanoTest uses sensors with only one electrode. A conductive disc mounted on the rear of the indentation tip serves as the second electrode.
####
About PI (Physik Instrumente)
About Micro Materials
A particularly flexible nano indentation system originates from Micro Materials Ltd. based in Wrexham/UK. The Nano-Test (Fig. 2) is used as an informative testing system in various fields world-wide. It is not only used in research and development but also in production and the testing of fi nished products.The current MML NanoTest system offers a range of methods for materials characterization, including nanoindentation, nano-impact and nano-scratch and wear measurements. It is in use in leading universities and industrial laboratories all over the world.
For more information on the NanoTest system visit www.micromaterials.co.uk
For more information on piezo nanopositioning systems and capacitive sensors visit www.pi.ws
or www.nanopositioning.net
For more information, please click here
Contacts:
Stefan Vorndran
VP Marketing
-----------------------------
PI (Physik Instrumente) L.P.
16 Albert St.
Auburn, MA 01501
email:
Tel: 508-832-3456,
Fax: 508-832-0506
http://www.pi.ws
http://www.pi-usa.us
Copyright © PI (Physik Instrumente)
If you have a comment, please Contact us.Issuers of news releases, not 7th Wave, Inc. or Nanotechnology Now, are solely responsible for the accuracy of the content.
Related News Press |
News and information
Beyond wires: Bubble technology powers next-generation electronics:New laser-based bubble printing technique creates ultra-flexible liquid metal circuits November 8th, 2024
Nanoparticle bursts over the Amazon rainforest: Rainfall induces bursts of natural nanoparticles that can form clouds and further precipitation over the Amazon rainforest November 8th, 2024
Nanotechnology: Flexible biosensors with modular design November 8th, 2024
Exosomes: A potential biomarker and therapeutic target in diabetic cardiomyopathy November 8th, 2024
Announcements
Nanotechnology: Flexible biosensors with modular design November 8th, 2024
Exosomes: A potential biomarker and therapeutic target in diabetic cardiomyopathy November 8th, 2024
Turning up the signal November 8th, 2024
Nanofibrous metal oxide semiconductor for sensory face November 8th, 2024
Tools
Turning up the signal November 8th, 2024
Quantum researchers cause controlled ‘wobble’ in the nucleus of a single atom September 13th, 2024
Faster than one pixel at a time – new imaging method for neutral atomic beam microscopes developed by Swansea researchers August 16th, 2024
The latest news from around the world, FREE | ||
Premium Products | ||
Only the news you want to read!
Learn More |
||
Full-service, expert consulting
Learn More |
||